Series ESPEC Platinous J it is the family of the temperature and humidity chambers, which consist of four types of chambers with volumes from 120 to 800 litres.
Seria J is an eighth generation of Platinous chambers – J stands for Japan Made, it is the shortest possible description of those chambers. Optimal construction, proven by customers, highest quality, reliability in working and latest technical solutions, they are basic features of mentioned chambers. Unique construction of  temperature regulation systems allowed for significantly decreasing the energy consumption – for some settings of parameters even 70% when comparing to chambers which are available in the market.

SEE SIMULATOR P300 CONTROLLER on ESPEC NA webpage.

Features

Features:

  • Internal capacities 120, 225, 408 (306) i 800 litres,
  • Available in versions: climatic test chamber (controlling of temperature and humidity) – types PR, PL i PSL or temperature test chamber (controlling of temperature) – types PU i PG,
  • Special versions for low humidities, series PDR/PDL – temperature range from -20 or -40 to +100 °C, humidity control range from 5 to 98 %RH,
  • Special versions for tests with higer temperatures and higher humidities, PHP series – temperature range from +10 to +100 °C, humidity range from 40 to 98 %RH,
  • Minimal temperatures: -20, -40 lub -70 °C,
  • Maximum temperatures: +100 °C, possibility of extending to +150 °C or +180 °C,
  • Temperature change rate according to PN-EN 60068-3-5: 3 °C/min for heating, 2 °C/min for cooling,
  • Relative humidity control range from 20 to 98 %RH,
  • Test chamber and housing made from stainless steel,
  • Air colled refrigeration system,
  • High precision of regulating temperature / humidity,
  • Innovative solutions for refrigeration system which allows to decrease of power consumption by the chamber.

N-Instrumentation:

  • Controller euquipped with touch screen TFT display 16,5 cm,
  • 10 pre-programed tests i.e: IEC 68-2-30 i MIL-STD 810,
  • Possibility of saving up to 40 user tests, each with 99 steps,
  • Ethernet Interface – build -in web sever: Web Manager, possibility of programming, controlling and monitoring the chamber by the web browser – Web Controller,
  • Interface USB 2.0 – possibility of connecting flash type memory (pendrive) – saving parameters of the test, reading and saving programs, possibility of edit test programs using an external PC and re-transfering them back to the chamber. Back trace function: possibility of tracking chamber parameters in diagnostic purposes – when alarm occurs, the technician can remotely diagnose the chamber. Iincluded software: Pattern Manager Lite for PC which allows to making / editing / archiving / processing and saving the datas (for example, making graphs).
  • 10 pre-programed tests i.e: IEC 68-2-30 i MIL-STD 810,
  • Possibility of saving up to 40 user tests, each with 99 steps,
  • Ethernet Interface – build -in web sever: Web Manager, possibility of programming, controlling and monitoring the chamber by the web browser,
  • Interface USB 2.0 – possibility of connecting flash type memory (pendrive) – saving parameters of the test, reading and saving programs, possibility of edit test programs using an external PC and re-transfering them back to the chamber. Back trace function: possibility of tracking chamber parameters for diagnostic purposes – when alarm occurs, the technician can diagnose the chamber, included software: for PC which allows to making / editing / archiving / processing and saving the datas (for example, making graphs).

Included in the delivery:

  • 4 adjustable wheels,
  • Cable port 50 mm I.D in left side wall,
  • Tank for water for humidifying system,
  • Build in hour meter,
  • Overheat protector.
Temperature (and Humidity) test chamber Platinous J
Type Temperature range1,2 Humidity range
Internal volume
PR -20 to +100 °C 20 to 98 %RH 120, 225, 408 i 800 liters
PL -40 to +100 °C
PSL -70 to +100 °C 306 i 800 liters
PDR -20 to +100 °C 5 to 98 %RH 408 i 800 liters
PDL -40 to +100 °C 5 to 98 %RH
PHP Ambient
temperature +10 to +100 °C
40 to 98 %RH 219, 398 i 784 liters
PCR* -20 to +100 °C 30 to 90 %RH 312 liters

1 – optional +150 lub 180 °C for PR, PL, PU
2 – optional +150 lub 180 °C for PSL i PG typ 2 (306 liters)
* Clean chamber with HEPA filters, clean class 5 according ISO

Temperature chambers Platinous, seria J
Type Temperature range Internal capacity
PU -40 to +100 °C 120, 225, 408 i 800 liters
PG -70 to +100 °C 306 i 800 liters


Temperature and Humidity Control Range

PARAMETRY

The standard family Platinous J chambers contains three groups of test chambers which differs in minimum temperature: PR from -20 to 100 ° C, PL from -40 to 100 ° C, and PSL from -70 to 100 ° C. Optionally, for each type, the upper temperature can be modified to +150 or 180 ° C. Detailed data for all chambers are given in the tables below.

Temperature (and Humidity) test chambers Platinous PL-xJ
Temperature range -40 to +100 °C
Humidity range 20 to 98 %RH
Model PL-1J PL-2J PL-3J PL-4J
Capacity
[Litres]
120 225 408 800
Internal dimensions
SxWxG [mm]
500x600x400 500x750x600 600x850x800 1000x1000x800
Temp. and Humidity fluctuation* ±0.3 °C / ±2.5 %RH
Temperature variation in space ±1.5 °C
Temperature rate of change °C/min* 3 / 2
Allowable heat load [W] 850 1400 1500 2850
Inside dimensions
SxWxG [mm]
910x1440x873 910x1590x1073 1010x1690x1273 1410x1970x1273
Weight [kg] 270 340 420 610
Maximum current [A]
400 V / 50 Hz
9.4 10.4 10.4 21.0

 

Temperature (and Humidity) test chambers Platinous PR-xJ
Temperature range -20 to +100 °C
Humidity range 20 to 98 %RH
Model PR-1J PR-2J PR-3J PR-4J
Capacity
[Litres]
120 225 408 800
Internal dimensions
SxWxG [mm]
500x600x400 500x750x600 600x850x800 1000x1000x800
Temp. and Humidity fluctuation* ±0.3 °C / ±2.5 %RH
Temperature variation in space ±1.5 °C
Temperature rate of change °C/min* 3 / 2
Allowable heat load [W] 800 1100 1250
Inside dimensions
SxWxG [mm]
910x1440x873 910x1590x1073 1010x1690x1273 1410x1970x1273
Weight [kg] 260 305 365 480
Maximum current [A]
400 V / 50 Hz
8.0 9.5 9.5 19.5

 

Temperature (and Humidity) test chambers Platinous PSL-xJ
Temperature range -70 to +100 °C
Humidity range 20 to 98 %RH
Model PSL-2J PSL-4J
Capacity [Litres] 306 800
Internal dimensions
SxWxG [mm]
600x850x600 1000x1000x800
Temp. and Humidity fluctuation* ±0.3 °C / ±2.5 %RH
Temperature variation in space ±1.5 °C
Temperature rate of change °C/min* 5 / 2 5 /1
Allowable heat load [W] 700 2200
Inside dimensions
SxWxG [mm]
1010x1690x1273 1410x1983x1593
Weight [kg] 470 705
Maximum current [A]
400 V / 50 Hz
17.1 29.4

* Definitions according:
EN 60068-3-5:2008 and EN 60068-3-6:2009

Options

  • Door with viewing window,
  • Doors with large window,
  • Thermocouples. Sample thermocouple connections for measuring object temperature – enable controlling temperature of tested samples instead of air temperature,
  • Recorders T / RH recorder for temperature and humidity,
  • Automatic defrost system,
  • Internal glass door,
  • Additional stainless steel shelves,
  • Additional shelves with increased load capacity,
  • Strengthening of thge load capacity of the floor to 100, 200 or 300 kg (standard 70 kg),
  • Air flow speed regulation,
  • Baskets for small samples,
  • Additional cable ports in side walls and ceiling,
  • Communication interfaces RS-232, RS-485 or GPIB,
  • Additional internal chamber for reducing of the air flow,
  • Water-cooled refrigeration system,
  • Additional thermostats for protection against high and low temperatures,
  • Extending of the upper temperature range to +150 or +180 ° C.