Automatic Hydrostatic Head Tester FX 3000 HYDROTESTER IV

Producent

Grupa urządzeń

Zastosowanie

Automatic Hydrostatic Head Tester FX 3000 IV HYDROTESTER measures the resistance of plastic foils, coated and uncoated fabrics and non-wovens to water penetration by means of the dynamic, static and program test method. Your choice: manual, semi-automatic or – worldwide unique: – fully automatic operation without any operator attendance. There are three version of the apparatus: FX 3000-4L PUM with built-in pump and mechanical sample clamp for a pressure up to 500 mbar (5 m water column), FX 3000-4M with external compressed air supply and pneumatic sample clamp for a pressure up to 2,000 mbar (20 m water column) and FX 3000-4H with external compressed air supply and pneumatic sample clamp for a pressure up to 5,000 mbar (50 m water column).

Features

  • Measuring range: to 5000, 2000 or 500 mbar – depends of the instrument type,
  • Measuring accuracy: +/- 0.5 %  of displayed value +/- 1 mbar,
  • Pressure gradient (Dynamic Test Method): 1…1000 mbar/min.,
  • Test pressure (Static Test Method): 1 mbar … max. test pressure,
  • Maximum test duration for Static Test Method: to 7 days,
  • Program Test Method: Freely programmable time-pressure-sequences,
  • Touch screen display for digital and graphic display of test result and test progress.,
  • Clamping of speciemen: Manual clamping arm Models 4M und 4H supported by pneumatic cylinder,
  • Test area: standard 100 cm2, optional: 10, 26 lub 28 cm2,
  • Maximum sample thickness: 5 mm,
  • Space behind test head: 15 cm,
  • Calibration certificate,
  • Capacity of the water tank: 0,8 litra,
  • Data ports: USB, Ethernet i WiFi, communication with the PC,
  • Dimensions: 40 x 65 x 71 cm,
  • Compressed Air -4H i -4M: 0,4 do 0,8 MPa
  • Power Supply 230 V / 50 Hz, 200 W.

Options

  • Different measuring heads,
  • Automatic drop detection with optional Drop Detector eliminates the attendance time for all tests.,
  • Special envelopes for preventing penetration of water through the sample edges,
  • Optional integrated strip printer for printing of a test report,
  • Optional integrated software for evaluation and storage of the test results in XML format.

Standards

EN 1734:2000

EN 20811:1997

AATCC 127, ASTM F 903-C, ASTM F 1670, ASTM F 1671, BS 2823, DIN 53886, EDANA 120.2-02, EN 1734, EN 20811, INDA IST 80.4, ISO 811, ISO 1420 A, JIS L 1092 A, JIS L 1092 B-b, NF G07-057, SNV 198571