Global-N Temperature Cycling Chambers
Producent | |
---|---|
Grupa urządzeń | |
Zastosowanie |
The Global N chambers manufactured by the ESPEC Company was designed for making tests in which the fast changes of the temperature is important. This kind of demands is most important parameter in automotive and electronic industry.
Features
- Four interior volumes: 380, 470, 800, and 1000 liters,
- Minimal temperature -40 or -70 °C,
- Maximum temperature +180 °C,
- Relative Humidity control range from 10 to 95 %RH,
- Possibility of changing temperature with speeds from 5 K/min to 20 K/min depends of the model,
- Watercolled or Aircolled refrigeration system,,
- P-300 touch-screen programmer with USB port,
- Possibility of controlling temperature by using Specimen Temperature Measurement Sensor.
Options
- Control of parameters in the test area by using Specimen Temperature Measurement Sensor,
- Web Controller – full control of tests by using an internal network or Internet, sending informations about the stasus of the chamber by an e-mail,
- System which allows to work in very low humidities,
- System of shock-type of cooling by using a liquid Nitrogen,
- Additional shelves and Cable ports.
TYPES
Types and volumes:
Chambers with internal volume: 380 liters | |||||
Type | Capacity | Temperature control range | Humidity control range | Heat up rate | Pull down rate |
EGNX12-4CWL/4CAL | 380 L | -70 to +180 °C | 10 to 95 %RH | 7 K/min | 6 K/min |
EGNX12-6CWL/6CAL | 15 K/min | 11 K/min | |||
EGNX12-7.5CWL | 15 K/min | 15 K/min | |||
EGNL12-4CWL/4CAL |
-40 to +180 °C
|
10 K/min | 8 K/min | ||
EGNL12-6CWL/6CAL | 15 K/min | 12 K/min |
Chambers with internal volume: 800 liters | |||||
Typ | Capacity | Temperature control range | Humidity control rate | Heat up rate | Pull down rate |
EGNX28-4CWL/4CAL | 800 L | -70 to +180 °C | 10 do 95 %RH | 10 K/min | 2.75 K/min |
EGNX28-6CWL/6CAL | 10 K/min | 5.5 K/min | |||
EGNX28-12CWL | 15 K/min | 12.5 K/min | |||
EGNX28-15CW | 20 K/min | 20 K/min | |||
EGNL28-4CWL/4CAL |
-40 to +180 °C
|
10 K/min | 4 K/min | ||
EGNL28-6CWL/6CAL | 10 K/min | 8 K/min | |||
EGNL28-7.5CWL | 20 K/min | 12 K/min | |||
EGNL28-12CWL | 20 K/min | 15 K/min |
NOTE. For all models, there are available chambers without controll of humidity (temperature chambers).